Brookeld Engineering Labs., Inc. Page 37 Manual No. M03-165-F0612
TORQUE = 24.2%
MIN: 15 SEC: 13
Figure IV-26
Note: When this mode has begun, a press of the MOTOR ON/OFF/
ESCAPE key will cancel the time to torque sequence and return
the user to the screen of Figure IV-14.
Thesecondsdisplaywillincrementfromzero(0)tofty-nine(59)inone(1)secondintervalsand
the current value of the viscometer torque will be updated continuously. When seconds reaches
fty-nine(59),theminutesvaluewillincrementbyone(1)minute.Thiswillcontinueuntiltheuser
selectedtorquevalueisattainedatwhichpointtheviscometerwilldisplaythefollowingscreen:
22MIN 54SEC: 85%
TIMED TORQ DONE
Figure IV-27
At this point the viscometer will stop the motor and continue to display this screen until any key
(except the UP or DOWN ARROW keys, the PRINT key or SELECT DISPLAY key) is pressed. The user can,
while this display is current, press the UP or DOWN ARROW keys to view the viscosity that was
currentatthetimethedisplaywasfrozen.Thedisplaywouldappearasfollows:
cP 123.5e6
TIMED TORQ DONE
Figure IV-28
The display will switch between that of Figures IV-27 and IV-28 for each press of the UP or DOWN
ARROW keys. As stated above, pressing any key (except the UP or DOWN ARROW or PRINT keys)
will cause the viscometer to exit the Time To Torque mode and resume operation with the screen
of Figure IV-23 displayed awaiting another Timed Torque run.
The user can press the PRINT key while in either of these two screens (Figures IV-26 and IV-27) to
send one standard print string to the attached printer as many times as the user presses the PRINT key.
In addition, the PRINT key can be pressed during the actual measurement to obtain instantaneous
data. Pressing any other key will exit this mode and return the viscometer to normal operation.
Note: For both of the methods of Sections IV.3.1 and IV.3.2 the following
apply:
1. For the Timed Stop method, the DV-II+Pro viscometers will retain the last value for the
time interval so that it will become the default the next time the user elects to use this
method.
2. For the Time To Torque method, the DV-II+Pro viscometers will retain the last entered
torque value for use when next the user elects to perform a time to torque test.
3. The user can set up a desired print interval time, then set the viscometer to the continuous