Brookfield DV2+Pro Viscometer Bedienungsanleitung Seite 18

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 101
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 17
Brookeld Engineering Labs., Inc. Page 18 Manual No. M03-165-F0612
SR 40.0 20.1C
6.0RPM % 15.6
Figure II-13
OnemorepressoftheSELECTDISPLAYkeywouldresultinareturntotheviscosityscreen,
as shown in Figure II-11.
Note:
1. Youmaystepthroughthedisplayatanytime.Thiswillnotinterrupt
any Viscometer calculations that are in progress.
2. Display of shear rate and shear stress requires selection of appropriate
spindles. Otherwise, values displayed will be zero (0). Refer to
Appendix D.
Shear rate and shear stress values will be displayed for any spindle with a SRC value greater
than zero.
Units of Measurement
TheDV-II+ProViscometercanbeconguredusingtheSETUPoption(SectionIV.2.2)todisplay/
print in either the CGS or SI system of units.
II.5
ENTER
AUTO
RANGE
Autorange
The
ENTER/AUTO RANGE key functions as Auto Range and allows you to determine the maximum
calculated viscosity (full scale reading) possible with the current spindle/speed setting only when in
the default screen. Pressing the key at any time will cause the current viscosity display to change
and show that maximum viscosity. Thescreenareadisplaying%(torque)willnowdisplayaashing
“%100.0”toindicatethisspecialcondition.Thismaximumviscosityandashing%100.0value
will be displayed for as long as the
ENTER/AUTO RANGE key is depressed. Figure II-14 shows the
AUTO RANGE function for the situation where the No. 1 LV spindle is rotating at 60 rpm. The full
scale range is 100.0 cP (or 100.0 mPa
.
s).
cP 100.0 S61
60 RPM % 100
Figure II-14
Note:
1. If the RPM is 0.0, the maximum viscosity displayed will be 0.0 cP (or 0.0 mPa
.
s).
2. While the Viscometer is in the Auto Range mode, any data sent to an attached
printerorcomputerreectsthedisplayedvalues(i.e.AutoRangevalues).
3. This function is only available when in the default screen.
4. If the motor is Off, Auto Range is not available.
Seitenansicht 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 100 101

Kommentare zu diesen Handbüchern

Keine Kommentare