Brookeld Engineering Labs., Inc. Page 55 Manual No. M13-2100
TIP:TorqueontheDV3Tisequivalenttothe%readingfromtheDV-IIIUltraandtheYR-1
Rheometer.
StressisthecalculatedstressbasedontheSpeed,Torque,andSpindle.
StrainistheApparentStrainplacedonthesample.TheApparentStrainistheangulardistance
thatthespindlerotatesinthesample.
TemperatureistheinputvaluefromaconnectedBrookeldtemperatureprobe.TheDV3Tis
providedwithaDVP-94Yprobethatcanbeinsertedintothetestsampleorawaterbath.The
Cone/PlateversionoftheDV3Tcanbeutilizedwithasamplecupthatincludesanembedded
temperatureprobe.SomeBrookeldaccessoriesincludetemperatureprobes(Thermosel)or
optionalchamberswithembeddedtemperatureprobes(SmallSampleAdapter).Theunitsof
temperaturearedenedintheGlobalSettingssectionoftheSettingsMenu
.
Thetemperaturedisplaywillbe----whennotemperatureprobeisconnected.
SpeedistheselectedspeedfromtheCongureYieldTestscreen.
Thestatusofthetestcongurationisindicatedinthegreybarabovethedataelds.Aconguration
thathasbeensavedwillbeindicatedbyadisplayofthetestlename.Acongurationthatisnot
savedwillbeindicatedbyadisplayofUnsavedTest.
TheReal Time GraphisdisplayedatthebottomoftheRunningYieldTest.Thisgraphshowsall
measurementsduringthelifeofthetest.TheY-Axiscanrepresentanydataeld.Thedataeld
representedonthegraphisshownbythebluedot;forexample,inFigureIV-7,Torqueisshown
onthegraph.Anydataeldcanbeselectedbytouchingthescreen.
TheRealTimeGraphwillindicatedatapointscollectedaspartoftheTestwithabluedotindicating
thenalpoint.Attheconclusionofthetest,theRealTimeGraphisdisplayedforreview.Once
theOKbuttonispressed,thescreenwilltransitiontoResults.
QCLimitswillbeindicatedontheRealTimeGraphwhentheselectionforY-Axisisthesameas
theselectionforQCLimits(seeSectionIII.9).
TwoCommandKeysareavailableontheRunningYieldTestScreen:StopTestandViewTest.
StopTest:Immediatelystopthecurrenttest.Theuserwillbepresentedwithaconrmation
box.ThetestwillcontinuetorununtilYesisselectedintheConrmationbox.Theuserwill
bepresentedwiththeResultsscreenwithanydatathatwascollected.
ViewTest:Changetheviewfrommeasurementtothetestparameters.TheViewTestscreen
includesalivedisplayoftemperature.ViewTestincludestheStopTestcommandkeyanda
BackArrowintheNavigationBar.TheBackArrowisusedtoreturnthedisplaytotheRunning
YieldTestview.