Brookeld Engineering Labs., Inc. Page 75 Manual No. M13-167-B0614
Calibration Procedure for a Small Sample Adapter
Brookeldrecommendsatwostepcheck.First,verifythecalibrationoftheviscometer
usingthestandard viscometer spindles(LV #1-3,RV#2-6, HA#2-6andHB #2-6or
cone/platespindles)asdetailedinthisappendix.Second,verifythecalibrationofthe
viscometer using the Small SampleAdapter. The use of an accessory device may
increasetheaccuracyofmeasurementassociatedwiththeDV2T.
WhenaSmallSampleAdapterisused,thewaterjacketisconnectedtothewaterbath
andthewaterisstabilizedatthepropertemperature:
1) Puttheproperamountofviscositystandarduidintothesamplechamber.The
amountvarieswitheachspindle/chambercombination(refertotheSmallSample
Adapterinstructionmanual).
2) Placethesamplechamberintothewaterjacket.
3) Putthespindleintothetestuidandattachtheextensionlink,couplingnutand
freehangingspindle(ordirectlyattachthesolidshaftspindle)totheDV2T.
4) Allow30minutesforthe viscositystandard,samplechamberandspindle to
reachtesttemperature.
5) MeasuretheviscosityandrecordtheViscometerreading.
Note: The spindle must
rotate at least ve (5) times before readings are taken.
Calibration Procedure for a Thermosel System
Brookeldrecommendsatwostepcheck.Firstverifythecalibrationoftheviscometer
usingthestandard viscometer spindles(LV #1-3,RV#2-6, HA#2-6andHB #2-6or
cone/platespindles)asdetailedinthisappendix.Secondverifythecalibrationofthe
viscometerusingtheThermosel. Theuseofanaccessorydevicemayincreasethe
accuracyofmeasurementassociatedwiththeDV2T.
WhenaThermoselSystemisused,thecontrollerstabilizestheThermoContainerat
thetesttemperature.
1) Put the proper amount of HT viscosity standard uid into the HT-2 sample
chamber. The amount varies with the spindle used (refer to the Thermosel
instructionmanual).
2) PlacethesamplechamberintotheThermoContainer.
3) Putthespindleintothetestuidandattachtheextensionlink,couplingnutand
freehangingspindle(ordirectlyattachthesolidshaftspindle)totheDV2T.
4) Allow30minutesforthe viscositystandard,samplechamberandspindle to
reachtesttemperature.
5) MeasuretheviscosityandrecordtheViscometerreading.
Note: The spindle must
rotate at least ve (5) times before readings are taken.
Calibration Procedure using UL or DIN Adapters
Brookeldrecommendsatwostepcheck.First,verifythecalibrationoftheviscometer
usingthestandard viscometer spindles(LV #1-3,RV#2-6, HA#2-6andHB #2-6or
cone/platespindles)asdetailedinthisappendix.Second,verifythecalibrationofthe
viscometerusingtheULorDINAdapters.Theuseofanaccessorydevicemayincrease
theaccuracyofmeasurementassociatedwiththeDV2T.