Brookeld Engineering Labs., Inc. Page 44 Manual No. M13-167-B0614
Change Password: Aspecicusercanchangetheirownpasswordatanytime.The
usermustrstenterthecurrentpassword.Thentheuserwillbe
promptedtoenterandconrmthenewpassword.
TIP:Iftheadministratorpasswordislost,itcanbereset.Please
contactBrookeldoryourBrookeldrepresentative.Remember
tohavetheinformationfoundintheAboutscreenavailable
(seeSectionIV.1).
IV.3 Global Settings
TheGlobalSettingsmenuincludessettingsthataffecttheoperationoftheDV2Tacross
allusersandmenus.Thesesettingsareindependentofuserlogin.TheveGlobal
Settingsinclude:
Measurement Units - Selecttheunitofmeasureforseveralparameters.
RegionalSettings - SpecifyLanguage,number,time,anddateformats.
GlobalAlarm - Select a single measurement parameter for alarm status.
Spindle List - Congurethedisplayofavailablespindles;createaspecial
spindle.
Speed List - Congurethedisplayofavailablespeeds;createanewspeed.
Measurement Units: TheMeasurementUnitsmenudisplaysthecurrentselectionsfor
measurementunits.Eachunitcanbechangedbypressingtheblue
downarrowandselectingthedesiredvalue.MeasurementUnits
arepartofthedataleforsavedresults.Availableunitsinclude:
Viscosity cP centipoise
P Poise
mPa
•
s milliPascal seconds
Pa
•
s Pascal seconds
cSt centistokes
mm
2
/s millimeterssquaredpersecond
1000cP=10 P=1000 mPa
•
s=1Pa
•
s
1cSt=1mm
2
/s
TIP: When cSt units are chosen for viscosity, a
densityvaluemustbeenteredintheDensity
eldinCongureViscosityTest.
Speed RPM revolutions per minute
1/s reciprocal seconds
TherelationshipbetweenRPMand1/sisdenedby
theSCRvalueoftheselectedspindle.SeeAppendix
D for details.
It's good information