Brookeld Engineering Labs., Inc. Page 8 Manual No. M/05-370-B0507
II. QUICK START
1. Unpack the instrument according to Section I.5.
2. Attach the selected probe. See Section IV.2 for more information.
3. Placethesampleonthebasetable.Adjustthetableheightsothatthesurfaceofthe
sample is within 5 mm of the probe (see Section III.3).
4. Set the test mode to Normal See Section III.6 for more information.
5. Setthetriggervalueto0.5%oftheloadcellrange.Forexample,use5gfora1000gload
cell instrument.
6. Set the test speed and distance. See Section IV.3 for more information.
7. Press the START button.
8. Record the Peak Load and/or Final Load in grams.
9. Clean the probe and remove sample.