Brookeld Engineering Laboratories, Inc. Page 32 Manual No. M/98-211-B0104
Brookeld Engineering Laboratories, Inc. Page 32 Manual No. M98-211-E0912
IV.2 DV-III Speed/Time Pair Programs for Making Viscosity Measurements
This programming method allows the operator to control the DV-III Ultra through the variables of
speed and time. These speed/time pairs instruct the rheometer to operate at a speed of rotation for
a certain period of time. Programs can be created with up to 25 steps. The DV-III Ultra can store
up to 10 programs. Upon completion of a program, the data may be viewed on the DV-III Ultra
display, analyzed or printed to an attached parallel or serial printer.
Two examples of programs are shown below:
Collect Data Over Time Collect Data At Several Speeds
Step RPM Time Step RPM Time
1 100 00:12 1 2.5 01:00
2 100 00:12 2 5 00:30
3 100 00:12 3 10 00:30
4 100 00:12 4 20 00:15
5 100 00:12 5 50 00:15
Five viscosity data points Five viscosity data points will be
will be collected over collected at ve speeds over
one minute. 150 seconds.
This program mode is accessed by pressing the program key and selecting number 1; 1 = DV-III.
The creation, editing and execution of DV-III programs are described in the following sections.
There are two types of test programs:
1) Next Speed Set where the test speeds are programmed, and the operator must signal the DV-
III Ultra to change speeds (and therefore take a reading) by pressing the ENTER key.
2) Prog Speed Set where the DV-III Ultra will perform the test automatically.
Each step of a program has two variables - speed and hold time. The reading is taken at the end of the
hold time interval in a Prog Speed Set or when the ENTER key is pressed in a Next Speed Set.
If the rst step hold time interval is 0 seconds, the program is a Next Speed Set type. If the rst
step interval is 1 second or more, the program is a Prog Speed Set type.
S SPEED SET SELECTION AND PROGRAMMING
The DV-III Ultra viscometer allows for the retention of a maximum of 10 speed sets with up to
25 discrete speeds per speed set. The program locations are numbered 0 through 9. These speed
sets are retained in EEPROM memory for those times when the DV-III Ultra is not powered up.
To access a previously programmed speed set or to enter data for a new speed set, the user presses
the 1” key when in the display of Figure IV-1 and is presented with the screen shown in Figure
IV-2:
SPEED SET OPERATIONS
1 = ENTER/EDIT A PGM
2 = CLEAR SPEED SET
3 = USE SPEED SET
Figure IV-2