20. Gently slide the torsion spring through the aluminum clamp sleeve and into the
dial. Make sure the bottom of the torsion spring is seated properly inside the dial.
Tighten the set screw located at the top of the dial to secure the torsion spring.
21. Replace the stop block and the two screws. Make sure that the zero dial reading
is aligned with the ll on the stop block.
22. Press down on the torsion spring until it is ush with the Aluminum Clamp
Sleeve and tighten the set screw.
23. Replace the bob and sleeve.
24. Re-attach the main housing cover and re-calibrate.