Brookfield Engineering Labs., Inc. Page 1 Manual No. M/97-241-B1101BROOKFIELD KU-1+ VISCOMETEROperating InstructionsManual No. M/97-241-B1101SPECIALIS
Brookfield Engineering Labs., Inc. Page 10 Manual No. M/97-241-B11015) Total allowable error for the calibration check in this example is 104.8 KU, ±
Brookfield Engineering Labs., Inc. Page 11 Manual No. M/97-241-B1101590 119.0 660 122.8 730 126.8 800 131.2 870 134.7 940 137.6 1010 139.9 1080 141.05
Brookfield Engineering Labs., Inc. Page 12 Manual No. M/97-241-B1101 Appendix B - ReferencesThe KU-1+ Viscometer is compatible with:ASTM D 562 Standar
Brookfield Engineering Labs., Inc. Page 13 Manual No. M/97-241-B1101 Appendix C - Warranty Repair and ServiceWarrantyBrookfield Viscometers are guaran
Brookfield Engineering Labs., Inc. Page 2 Manual No. M/97-241-B1101TABLE OF CONTENTS I. INTRODUCTION ...
Brookfield Engineering Labs., Inc. Page 3 Manual No. M/97-241-B1101 I. INTRODUCTIONThe Brookfield KU-1+ Viscometer measures fluid viscosity in Krebs u
Brookfield Engineering Labs., Inc. Page 4 Manual No. M/97-241-B1101I.2 OptionsI.2.1 Air PurgeThe air purge option (Brookfield Part No.KU-1A) allows th
Brookfield Engineering Labs., Inc. Page 5 Manual No. M/97-241-B1101I.4 Can SpecificationsContainer DimensionsCan Height Bottom Flange DiameterU.S. Pin
Brookfield Engineering Labs., Inc. Page 6 Manual No. M/97-241-B1101 II. OPERATIONII.1 Set-Up1) Be sure that the power switch is off. Attach the power
Brookfield Engineering Labs., Inc. Page 7 Manual No. M/97-241-B1101has been put into the container, the surface of the fluid will be at the immersion
Brookfield Engineering Labs., Inc. Page 8 Manual No. M/97-241-B1101 Appendix A - KU-1+ Calibration InformationThe accuracy of the KU-1+ is verified us
Brookfield Engineering Labs., Inc. Page 9 Manual No. M/97-241-B11016) After you have recorded results in KU, then move the Selector knob to the "