Brookfield Engineering Labs., Inc. Page 29 Manual No. M/05-370-A1205
Appendix B - Troubleshooting
If no power to the instrument:
Check:
User replaceable fuses: two (2) fuses, 4 amp, 5 x 20mm, time-lag
Fuses located within power entry module on the rear of the instrument.
If fuses blow continually, serious damage to the instrument could result. Contact repair
services immediately.
Probe won’t attach:
Check thread on probe and probe shaft - remove any dirt
Base table won’t move:
Ensure locking knob is loose
Ensure fine adjust nut is loose
Base table will not travel to the lowest position:
Loosen fine adjust nut and rotate base table
Instrument is unsteady:
Ensure that bench top is stable
Adjust the LFRA feet
Probe doesn’t move when start is pressed:
Ensure emergency stop button is not pressed in (rotate clockwise)
Ensure test mode is not Static Load
Ensure probe is not in contact with sample (or base table)